Thermoluminescence (TL) technology is commonly used in medicine, such as in radiation therapy for dose measurement on the skin and within the body to assess treatment and ensure that safe radiation levels are maintained. The Thermo Scientific™ HARSHAW TLD™ Model 3500 Manual Reader provides manual readout of TL chips, disks, rods and cubes in a broad variety of sizes. This reader is suitable for all kinds of applications that require thermoluminescence dosimetry (TLD) detectors. Due to planchet heating, this versatile model is able to read powders, an essential feature for food testing and analysis.
The HARSHAW TLD Model 3500 Manual Reader includes a sample drawer for a single element TL detector, a linear, programmable heating system, and a cooled photomultiplier tube with associated electronics to measure the TL light output. Planchet heating incorporates a welded thermocouple for best temperature reproducibility. The operational software, Thermo Scientific™ WinREMS™, which runs on a separate computer, provides the user interface, the reader control and the applications software.
Ambient Light Exposure | Functions within specifications when exposed to ambient light up to 1000 Lux with cover on the reader |
Dose Range | Measurement range: 10 μGy to 1 Gy (linear), 1 Gy to 20 Gy (supralinear) Detection threshold: <10 μGy based on 2.26 standard deviations for 10 sequential readings of an unexposed dosimeter |
Dosimeter Reusability | Typically>500 |
Dynamic Range | 7 decades |
Flow Rate | 130 liters/h, subject to external regulation |
PMTs | Thermoelectric PMT cooler for Maximum gain stability |
Power Supply | Selectable, 110/120 VAC ±10%, 60 Hz or 220/240 VAC ±10%, 50 Hz |
Sensitivity | Dependent on config. Used |
Time and Temperature Profiles | Heating temperature capability up 600°C; Typical cycle time per detector: 30 seconds |
Batch Uniformity | ±30% from the factory standard |
For Use With | Thermo Scientific TLD Card Dosimeters, Thermo Scientific TLD Materials, Thermo Scientific TLD Accessories |
Heating Method | Multiple, programmable, linear heating cycles include pre-read anneal, acquire and post-read anneal cycles |
Humidity | Functions within specification after 24 hour |
Linearity | Better than 1% |
Operating Temperature | 15°C to 40°C |
Radiation Type | Photons, Electrons, Neutrons |
Reference Light | Temperature-stabilized LED |
Repeatability | Less than 2% variation based on one standard |
Stability | HV stability: ±0.005% per hour, ±0.02% per 8 hours |
Storage Temperature | -10°C to +60°C |
Weight | 25 kg |
Dosimetry Performance Using LiF:Mg,Ti (TLD-100/600/700):
Instrument Performance:
Ordering Alerts:
Computer requirements: Any commercially available PC system with Windows™ 7, XP, 2000, or NT 4. 32 bit OS only. RS232 strongly recommended. (Please contact your sales representative for approved USB to RS232 converters).
Compatible with:
Thermo Scientific TLD Materials, Thermo Scientific TLD Accessories
Recommended for:
Treatment planning verification, diagnostic dose studies, quality assurance, whole body dosimetry, extremity dosimetry, environmental dosimetry, high dose applications (radiation hardness testing), irradiated food analysis, dating, and many more.
Request for Quote