Get near-tissue equivalent thermoluminescence dosimetry (TLD) performance with a high-capacity reader for all types of TLD elements and health physics applications. The Thermo Scientific™ HARSHAW TLD Model 5500 Automatic Reader provides unattended operation for up to 50 dosimeters and accommodates TL chips, disks, rods and cubes in a broad variety of sizes. This reader is suitable for all kinds of applications that require TLD detectors. Clinical applications are: Radiotherapy planning verification, total body irradiation dose verification, critical organ dose verification, diagnostic dose studies including CT, and many others where measurement quality assurance is critical.
The HARSHAW TLD Model 5500 Manual Reader includes a carrier disk for up to 50 TL detectors, a linear, programmable contactless nitrogen heating system, and a cooled photomultiplier tube with associated electronics to measure the TL light output. The operational software, Thermo Scientific™ WinREMS™, which runs on a separate computer, provides the user interface, the reader control and the applications software.
Ambient Light Exposure | Functions within specifications when exposed to ambient light up to 1000 Lux with cover on the reader |
Description | Model 5500 TLD Reader Instrument |
Detector Type | Thermoluminescence (TL) Detector |
Dose Range | Measurement range: 10 μGy to 1 Gy (linear), 1 Gy to 20 Gy (supralinear) Detection threshold: <10 μGy based on 2.26 standard deviations for 10 sequential readings of an unexposed dosimeter |
Dosimeter Reusability | >500 |
Dynamic Range | 7 decades |
Flow Rate | Flow Rate: internally regulated to 400 liters/h if specified pressure is provided |
PMTs | Thermoelectric PMT cooler for Maximum gain stability |
Power Supply | 100/120/220/240 VAC |
Product Type | Radiation Dosimeter Reader |
Sensitivity | Dependent on config. Used |
Time and Temperature Profiles | Multiple, programmable, linear heating cycles include pre-read anneal, acquire and post-read anneal cycles; Typical cycle time per detector: typically 30 seconds |
Batch Uniformity | ±30% from the factory standard |
For Use With | Thermo Scientific TLD Card Dosimeters, Thermo Scientific TLD Materials, Thermo Scientific TLD Accessories |
Humidity | Functions within specifications after 24-hour exposure to 95% and subsequent 6-hour recovery |
Linearity | Better than 1% |
Operating Temperature | 15°C to 40°C |
Radiation Type | Photons, Electrons, Neutrons |
Reference Light | Temperature-stabilized LED |
Repeatability | for 1 mGy (100 mrad) 137 Cs doses, >2% standard deviation of 10 sequential measurements |
Stability | HV stability: ±0.005% per hour, ±0.02% per 8 hours |
Storage Temperatur | -10°C to +60°C |
Weight | 40 kg |
Dosimetry Performance Using LiF:Mg,Ti (TLD-100/600/700):
Instrument Performance:
Ordering Alerts:
Computer requirements: Any commercially available PC system with Windows™ 7, XP, 2000, or NT 4. 32 bit OS only. RS232 strongly recommended. (Please contact your sales representative for approved USB to RS232 converters).
Compatible with:
Thermo Scientific TLD Materials, Thermo Scientific TLD Accessories
Recommended for:
Treatment planning verification, diagnostic dose studies, quality assurance, whole body dosimetry, extremity dosimetry, environmental dosimetry, high dose applications (radiation hardness testing), and many more.
Request for Quote